OI Analytical FS3700 Continuous Flow Analyzer

Micromac C Colorimetric Chemistry Analyser

Automates Wet Chemistry Test Procedures

  • Can perform both Segmented Flow Analysis (SFA) and Flow Injection Analysis (FIA)
  • Validates Hardware Set Up and Performance for every method supplied
  • Multiple FS3700’s can be linked to provide additional channels for simultaneous analysis.
  • Modular, Flexible Hardware

Versatile

Featuring an unmatched modular design, the FS3700 allows for flexible and versatile functionality to help meet your lab’s goals. Run different flow methods in different channels in tandem on the FS3700. Add pre-configured chemistry cartridges and perform validated analysis methods. Link Multiple systems together to provide additional analytical channels.

Easy Validation

Hardware configurations and performance of each supplied method are validated to provide users with a complete analysis solution. Validated methods for aqueous samples, soil, and plant extracts that support environmental compliance monitoring, process optimization, and research applications are available in easy-to-use, pre-configured chemistry cartridges.

Simple-to-Learn Interface

FlowView™ Software is unmatched in providing a user-friendly, icon based interface that simplifies scheduling, operation, and report generation from the FS3700. Compatible with Windows 32- or 64- bit operating systems, it requires minimal training and makes exporting and retrieving data a breeze.

FlowView™ features:

  • On-the-fly sequence editing, and calibration monitoring
  • LIMS-compatible import + export with customizable report generation
  • Refined algorithms for peak detection, baseline handling, and carry-over correction
  • System configuration and method parameters are archived with data in an individual result file
  • Unparalleled access to instrument component diagnostics in real-time, even during data collection

Methods*

Analyte Technique Method Operating Range MDL Throughput
Ammonia Nitrogen FIA USEPA Method

350.1

0.01-20 mg/L
ammonia as nitrogen
0.002 mg/L 30 samples
per hour
Available Cyanide FIA USEPA Method

OIA-1677-09

2.0 μg/L–5.0 mg/L 0.5 μg/L 30 samples
per hour
Nitrate/Nitrite FIA

SFA

USEPA Method

353.2

0.01–10.0 ppm
10.0-10,000 ppb0.005-10.0 ppm
0.001 ppm
1.0 ppb0.001 ppm
60 samples
per hour40 samples
per hour
Nitrogen

(Total Dissolved)

SFA USEPA Method
353.2
0.01 – 20 mg/L 0.006 mg/L N
as NO3
20 samples
per hour
Phenol

In-Line Distillation

SFA USEPA Method
420.4
1.0-500 ppb 0.5 ppb 22 samples
per hour
Phenol

(Post-Distillation)

FIA USEPA Method
420.4
0.01-2.00 ppm
10.0–2,000 ppb
0.002 ppm
2.0 ppb
90 samples
per hour
Phosphorus

(All Forms)

FIA USEPA Method
365.1
0.01–5.0 ppm
10.0-5,000 ppb
0.001 ppm
1.0 ppb
60 samples
per hour
Phosphorus

All Forms- Low Level

FIA USEPA Method
365.1
0.001-0.1 ppm
1.0-100 ppb
0.0003 ppm
0.3 ppb
45 samples
per hour
Phosporus(Total) FIA w/ In-line UV/Persulfate Standard Method
4500-P-I
0.010 mg/L – 10 mg/L P 0.002 mg/L 30 samples
per hour
Silica SFA Standard Method
4500-SiO2 E
0.02 – 20 mg/L SiO2 0.02 mg/L 60 samples
per hour
Sulfate FIA

Photometric

USEPA Method
375.2
1.0 mg/L –
25 mg/L
0.1 mg/L 40 samples
per hour
TKN

(Total Kjeldahl Nitrogen)

SFA, Gas Diffusion USEPA Method
351.2
0.01-20.0 ppm
10.0-20,000 ppb
0.001 ppm
1.0 ppb
40 samples
per hour

*More methods available, these are the most popular options

Technical Specifications

Analysis Module: 1 or 2 chemical analysis channels per chassis

Analysis Module Dimensions: 31 in. W x 17.5 in. D x 10.5 in. H; 78.74 cm W x 44.45 cm D x 26.67 cm H

FS 3700 Dimensions with 90-position Autosampler: Approximately 44 in. (112 cm) W

FS 3700 Dimensions with 360-position Autosampler: Approximately 51.5 in. (131 cm) W

Injection Valve: 8 or 10-port switching valve with chemically-inert wetted surfaces

Photometric Detector: 420-880 nm, with PEEK path lengths of 5-, 10- or 20-mm

Amperometric Detector: Silver working electrode; silver/silver chloride reference; stainless steel counter electrode

In-line Heater: Included as needed, mounted underneath chemistry cartridge user programmable in 1 ̊C increments

UV-digestion Module: Included as needed, mounted underneath chemistry cartridge

Peristaltic Pump: 24-channel, fits on top of analysis module

Autosampler: 90-position, X-Y-Z (90 samples + 9 standards); 360-position, X-Y-Z (360 samples + 10 standards)

Tubing: FEP Teflon® and EVA ethylene-vinyl acetate copolymer

Manifolds / Fittings: Polysulfone

Analysis Methods / Documentation: Validated chemistries for specific analytes / sample matrices with performance data

Operating Software: FlowView

Operating System: Windows® 7, Windows® 8, 8.1 and 10

Data Collection: 6 channels per instance of software; Multiple instances of software can be run on a single computer

PC to FS 3700 Communications: USB

Power Supply: 24VDC universal switching power supply for operation with 90-250VAC 50/60Hz source

Power Requirements: 110VAC/60 Hz or 230VAC/50 Hz

Weight (Analysis Module): 19.5 kg (43 lbs.), typical for analysis module and pump, two injection valves, chemistry cartridges, detector modules

Certifications: CE Safety EN 61010-1; EMC Immunity & Emissions EN 61326-1:2006

 

 

 

Resources