Micromac C
ON LINE ANALYZER FOR POTABLE, SURFACE, & WASTE WATER MONITORING
-
User-friendly interface
-
Fully Automated
- Low maintenance & operating cost
- Very low reagent consumption
- Easy to operate; fully documented plug in analyzer, no special training required
- Electronics and hydraulics completely separated
- RS232 bidirectional for remote control
- USB port for data download
- Yearly maintenance
ROBUST AND RELIABLE
Micromac C is designed for industrial and environmental on-line applications and features highly durable electronic, mechanical, and hydraulic components. The electronic and hydraulic components are completely separate to ensure long and reliable operation.
EASY TO INSTALL
The analyzer is delivered factory tested and ready to install,
including a set of spares for start-up operations.
AUTOMATIC CALIBRATION
Micromac C auto-calibrates to a factor within acceptance limits.
MEASUREMENT FREQUENCY
User selectable; between two measurements the analyzer remains
in stand-by mode, without reagents consumption.
OFF SCALE REANALYZE
Micromac C identifies off scale samples and reanalyze the sample
after automatic dilution
Technical Specifications
MEASURING PRINCIPLE:
- Colorimetric -dual beam
- ISE (Ion Selective Electrode)
- NDIR
- UV
MEASUREMENT TYPE: Batch – Multi parametric version: batch, sequential & multiwavelength
MEASURING FREQUENCY: Programmable
MEASURING TIME: From 5 minutes up to 60 minutes depending on the application
MEASURING POINTS Up to 6
OPERATOR INTERFACE: Membrane keypad & Graphic display or color graphic touch screen 8” (option)
OUTPUT SIGNAL AND COMMUNICATIONS PORTS: 4-20mA separated per each method or stream, galvanic insulation option; RS232 output ; RS485 optional; RTU Modbus optional
INPUT SIGNALS:
- Analysis: 1 digital contact with photocoupler, galvanically isolated
- Calibration: 1 digital contact with photocoupler, galvanically isolated
ALARM CONTACTS:
- High Limit : 1 potential free switch SPDT, max load 24 AC DC 0.5 A separated for each stream
- Calibration: 1 potential free switch SPDT, max load 24 AC DC 0.5 A
- General: 1 potential free switch SPDT, max load 24 AC DC 0.5 A, separated for each stream
ALARM MESSAGES: On operator interface
SAMPLE DELIVERY:
- Pressure: atmospheric
- Temperature: 10° – 35 °C
- Volume: 50/80 mL per analysis depending on analyzer model
- Connection: Standard silicone 2×4, other on request
- Waste: pressure free silicone 2×4 mm
REAGENT REPLACEMENT: From 3 to 5 weeks depending on the measurement frequency
ENVIRONMENTAL TEMP. 10-45°C (Reagents up to 25°C)
MOUNTING Wall mounting
STANDARD PROTECTION:IP 55, IP 65 optional on request
HARDWARE: PC104 standard microcontroller, integrated 8″ colour touch screen
POWER SUPPLY: 12 Vcc; external power supply from 110/220 Vac to 12 Vcc included in the scope of delivery
ABSORPTION: 4W stand by, 10 W analysis
WEIGHT: 30 Kg without reagents
DIMENSIONS: 800x420x280 mm (hxwxd)
SELF CLEANING FILTER
POWER SUPPLY: 12Vdc
SAMPLE PRESSURE/RATE: min 0.3 bar/30 l/h max 1 bar
COMPRESSED AIR FOR SELF CLEANING: max 2 bar
Applications for On Line Analyzers
Alkalinity (methyl orange)
Alluminum
Ammonia
Ammonia (fluorimetric)
Arsenic Total dissolved
Arsenic Total
Boron
Cadmium Total/Dissolved
Calcium
Chloride
Chlorine Free/Total
Chromium
Chromium Total
COD (Dichromate)
COD (Permanganate)
COD (UV 254nm)
Color
Copper
Copper Total
Cyanide Tot/Free (UV,Dist)
Cyanide Index
Ethylene glycol
Fluoride
Hardness
Hydrazine
Iron Total dissolved
Iron Total
Lead Total/Dissolved
Manganese
Manganese Total
Monochloram. & Total
Nickel
Nickel Total
Nitrate+Nitrite Hydrazine
Nitrate+Nitrite UV photored.
Nitrate (UV 220nm)
Nitrite
Nitrogen Total (Colorimetric)
Total Nitrogen (UV)
Phenol Volatile
Phenol Index
Ortophosphate
Silicates
Sucrose
Sulfide
TOC
Total Phosphorous
Zinc
Zinc Total