Micromac-1000

Portable Analyzer for Water Online Monitoring

Micromac-1000 is the only high sensitivity multiparametric portable analyzer in the market; wet chemistry methods normally used in the lab can now be run in the field! Measure any type of water sample with the wide range of EPA approved applications available, including high sensitivity methods for nutrients analysis in seawater.

  • Portable, easy to install & operate;
  • Low reagents consumption; only few hundreds microliters of reagents per analysis, low reagents and disposable cost;
  • Fully automated operation; no operator attending required;
  • Automatic calibration with calibration function.
  • Results data storage; the analyzer can store up to 400 measuring results including date, time and optical density.

Easy to use

Sample analysis with the Micromac-1000 is easy; an operator can run a sample by selecting a two key analysis function on the integrated keyboard.

Micromac-1000 can perform very complex analytical cycles; the results are displayed in concentration units and provided through the analog or the RS-232 port.

Always ready for analysis

At the end of each analytical cycle the analyzer remains in stand by mode, without reagent consumption. It’s always ready for next sample.

Easy to install

The factory-tested analyzer is delivered ready to be installed; start-up kit included. Simply prepare the reagents, connect sample and waste lines, and power up to begin analysis.

Sample dilution

The analysis can be run on either a whole sample or after dilution on operator request. Dilution mode will be automatically activated on over range samples, with a dilution factor, factory selected, up to 100.

Features and benefits

  • Portable, easy to install and to manage;
  • Low reagents consumption; only few hundreds microliters of reagents per analysis, low reagents and disposable cost;
  • Fully automated operation; no operator attending required;
  • Automatic calibration; it is enough to select the calibration function to perform a new calibration;
  • Results data storage; the analyzer can store up to 400 measuring results including date, time and optical density.

Technical Specifications

Measuring principle: Colorimetric

Detector: Multi-beam Colorimeter with Silicon Detector
Measurement type: Cyclic (Cyclic & Sequential MP version)
Measuring interval: Programmable
Measuring time: Dependent on the Analytical Method
Number of Parameters: up to 4
Number of Measuring Points: 1
Output Signals: RS-232 Serial Output
Input Signals: Digital Contacts with Optoisolated for Analysis and Calibration Start
Sample Delivery:
  • Pressure: Atmospheric
  • Temperature: 10° – 30 °C
  • Volume: 20/30 ml
  • Connections: Standard 3.2 x 1.6 mm or others on request

Operating Temperature: 4°-30 °C
Waste: Pressure Free
Mounting: Integrated Carrying Case
Protection: IP55
Reagents replacement: From 3 to 4 weeks, Dependent on Configured Methods
Reagent compartment cooling: Optional, Peltier Cell
Autonomy: Up to one month, Dependent on Measurement Interval Time
Hardware: PC-104 Industrial CPU; Integrated Keyboard and Graphic Matrix Display
Data Output: Standard RS 232 Serial Port; RS 485 Available on Request
Power supply: 12 Vdc, Provided through the Main Connecting Cable
Power absorption: 4 W in stand by, 10 W during analysis
Weight: 13 Kg without reagents
Dimensions: 550 x 350 x 110 cm (l x h x w)

Applications for Portable Analyzers

Alkalinity

Aluminum

Ammonia (colorimetric)

Ammonia(fluorimetric)

BOD (fluorimetric “Tryptophan like”)

Boron

Calcium

Chloride

Chlorine

Chlorine Total & Free

Chromium 6+

Copper

Cyanide Index

Ethylene Glycol

Hardness

Hydrazine

Iron Total Dissolved

Manganese

Monochloramine

Nickel

Nitrate (Hydrazine reduction)

Nitrate (VCl3 reduction)

Nitrate(UV reduction)

Nitrate (UV 220 nm direct reading)

Nitrite

Nitrogen Total

Phenol Index

Orthophosphate

Phosphorous Total

Silicates

Sulfide

TSS

Turbidity

Zinc

Resources